Product Name: Memo ETCO2 Capnograph ( Mini Capnograph)
Model No. : CMZ60MEMO
Technical Specification
Description | Compact, battery powered, quantitative capnograph for mainstream CO2 monitoring of adult, pediatric and infant patients.。 |
Measurements | CO2 partial pressure and respiratory rate |
Measuring principle | 2 channel NDIR type gas analyzer,No moving parts |
Warm up | Waveform displayed within 10 seconds, meets specifications within 2 minutes (at 25°C room temperature) |
Calibration | No routine calibration required |
CO2 Range | 0~99 mmHg
0~9.9 % 0~9.9 kPa |
CO2 Resolution | 1mmHg or 0.1kPa or 0.1% |
CO2 accuracy | 0~40mmHg ±2 mmHg
41~99mmHg ±8% of readings When RR is above 80 bpm ±12% of readings |
Drift of CO2 measurement
accuracy |
Short drift: Less than 1 mmHg offset in 4 hours
Long drift: Meet measurement accuracy requirements within 120 hours |
CO2 noise | Noise RMS less than 1mmHg at 5% CO2 |
Total system response time | Less than 500ms |
Recovery time after
defibrillator test |
Unaffected |
Respiratory rate | 3~150 bpm |
Respiratory rate accuracy | ±1 bpm |
ETCO2 Calculation Method | Peak of the expired CO2 waveform |
Compensation | Built-in atmospheric pressure sensor, automatic pressure compensation |
Display | 128 *128 pixels 1.44 inch TFT color display |
Dimensions | 51 x 43 x 45 mm |
Weight | <65 g (Included batteries) |
Mechanical robustness | Withstands repeated 1 m drops.
Meets the shock and vibration requirements for transport of EN ISO 80601-2-55:2011 clause 201.15.3.5.101.2 and EN 1789:2007 clause 6.3.4.2 and 6.4.1. |
Operating conditions | Temperature:0 – 40 ℃
Humidity:<90% (non-condensing) Atmospheric pressure:50-120 kPa |
Storage conditions | Temperature:-20 – 70 ℃
Humidity:<95% (non-condensing) Atmospheric pressure:50-120 kPa |